MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71992-01
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Information (3190)
Event Date 08/13/2020
Event Type  malfunction  
Manufacturer Narrative
No product has been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If the product is returned, the case will be re-opened and a physical investigation will be performed.The sensor is designed to report readings of 40 mg/dl to 500 mg/dl.It should be noted that this sensor does not give numeric value for readings less than 40 mg/dl.A "lo" display message indicates a reading less than 40 mg/dl.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
Customer reported receiving a low reading from their freestyle libre 2 sensor.Customer reported no trend arrows at the time of the call.Customer reported a low reading of "lo" (below 40 mg/dl) which was lower than lab reading of 200 mg/dl.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
 
Manufacturer Narrative
Sensor (b)(4) has been returned and investigated.The sensor plug is fully seated and no issues were observed.Extracted data from the returned sensor using approved software.The sensor was found to be in sensor state 5 (indicating normal termination).Inspected the plug assembly, no issues were observed.The current was applied to the sensor to perform linearity testing while in the test fixture.All results were within specification.No malfunction or product deficiency was identified.
 
Event Description
Customer reported receiving a low reading from their freestyle libre 2 sensor.Customer reported no trend arrows at the time of the call.Customer reported a low reading of "lo" (below 40 mg/dl) which was lower than lab reading of 200 mg/dl.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
MDR Report Key10792646
MDR Text Key214793632
Report Number2954323-2020-09920
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 12/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number71992-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2020
Initial Date Manufacturer Received 08/14/2020
Initial Date FDA Received11/05/2020
Supplement Dates Manufacturer Received12/16/2020
Supplement Dates FDA Received12/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-