Model Number 87047
Device Problems Material Deformation (2976); Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/18/2020
Event Type  malfunction  
Manufacturer Narrative
Event is reportable based on device analysis completed.Visual inspection revealed the sheath of the catheter was torn, exposing guide coil and wires were visible in the distal end.Dried saline was found in the luer, on the handle, shaft and tip.Dried body fluids was also noted on the handle.The steering knob and the tension control knob functioned properly on both lock and unlock positions.No abnormal resistance was felt when actuating the steering mechanism.Both right and left curves were 180 degrees out of plane.The complaint investigation conclusion code is design inadequate for purpose.
 
Event Description
Reportable based on device analysis completed on 10/08/2020.During an ablation procedure a intellanav mifi open-irrigated ablation catheter was selected for use.It was reported that the catheter shaft did not bend or curve.The catheter was removed from the patient and replaced with another catheter of the same model.The procedure was completed successfully without any patient complications.However, device analysis revealed torn catheter tubing on the distal end with exposing wires.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLANAV MIFI OPEN-IRRIGATED ABLATION CATHETER
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key10792886
MDR Text Key214809643
Report Number2134265-2020-15392
Device Sequence Number1
Product Code OAD
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/05/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/20/2022
Device Model Number87047
Device Catalogue Number87047
Device Lot Number0025075594
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/08/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/21/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-