• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problems Break (1069); Defective Component (2292)
Patient Problem No Patient Involvement (2645)
Event Date 10/12/2020
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) evaluated the iabp unit, and was able to reproduce the reported issue.To fix the issue, the fse replaced the fiber optic sensor assembly, performed all functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.Full event site name: (b)(6).
 
Event Description
It was reported that prior to use the cardiosave intra-aortic balloon pump (iabp) fiber optic port was broken.There was no patient involvement, and no adverse event reported.
 
Event Description
It was reported that prior to use the cardiosave intra-aortic balloon pump (iabp) fiber optic port was broken.There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
MDR Report Key10793463
MDR Text Key215618969
Report Number2249723-2020-01826
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/13/2020
Initial Date FDA Received11/05/2020
Supplement Dates Manufacturer Received12/15/2020
Supplement Dates FDA Received01/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-