Model Number 33422
Device Problem Fluid/Blood Leak (1250)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/14/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that the foley catheter had urine leak.
 
Event Description
It was reported that the foley catheter had urine leak.
 
Manufacturer Narrative
The reported event was inconclusive, as no sample returned for evaluation.A potential root cause for this failure mode could be due to "bad fit with the shaft." the device used for the treatment, though it was unknown whether the device related to the reported event or met specifications.The device was not returned for evaluation.The lot number was unknown, therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe stick in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD SILASTIC FOLEY CATHETER
Type of Device
SILASTIC FOLEY CATH 30CC
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key10793598
MDR Text Key215071007
Report Number1018233-2020-20749
Device Sequence Number1
Product Code EZL
UDI-Device Identifier00801741020018
UDI-Public(01)00801741020018
Combination Product (y/n)N
PMA/PMN Number
K951103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 02/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/05/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number33422
Device Catalogue Number33422
Was Device Available for Evaluation? No
Date Manufacturer Received02/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-