Catalog Number UNK ABSORB
Device Problem Patient-Device Incompatibility (2682)
Patient Problems Myocardial Infarction (1969); Thrombosis (2100)
Event Date 05/17/2019
Event Type  Injury  
Manufacturer Narrative
Event date is estimated.Implant date is estimated.The scaffold remains in the patient.The device was not returned for evaluation.A review of the lot history record and complaint history of the reported lot could not be conducted because the part and lot numbers were not provided.A conclusive cause for the reported wall apposition, patient effects, and the relationship to the product, if any, cannot be determined; however, the reported treatments appear to be related to circumstances of the procedure.The reported potential adverse events of myocardial infarction and thrombosis are listed in the absorb bioresorbable vascular scaffold system (bvss), instructions for use as known adverse events associated with the use of a coronary scaffold.There is no indication of a product quality issue with respect to manufacture, design or labeling.Literature attachment.Article title: very late scaffold thrombosis after everolimus-eluting bioresorbable scaffold implantation in patients with unremarkable interim surveillance angiography.The udi number is unknown as the part and lot #s were not provided.
 
Event Description
Patient 6.It was reported that this patient was featured in a journal article titled, "very late scaffold thrombosis [vlsct] after everolimus-eluting bioresorbable scaffold [brs] implantation in patients with unremarkable interim surveillance angiography".From the article: the patient was treated with two 3.5/28 mm brs for presentation with nstemi [non st elevation myocardial infarction].Follow-up angiography at 15 months after the index pci [percutaneous coronary intervention] showed a satisfactory result.866 days after the index pci, the patient presented with stemi caused by vlsct.St-segment elevation resolved after administration of heparin in the emergency department.There was timi 3 flow by coronary angiography.Oct [optical coherence tomography] showed thrombotic material surrounding malapposed struts in the lumen of the distal rca [right coronary artery] at the site of a focal scaffold discontinuity.A des [drug eluting stent] was implanted with a good procedural result.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSORB BIORESORBABLE VASCULAR SCAFFOLD SYSTEM
Type of Device
BIORESORBABLE SCAFFOLD
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key10796317
MDR Text Key216644147
Report Number2024168-2020-09286
Device Sequence Number1
Product Code PNY
Combination Product (y/n)Y
Reporter Country CodeGM
PMA/PMN Number
P150023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type literature
Reporter Occupation Physician
Type of Report Initial
Report Date 11/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK ABSORB
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/21/2020
Initial Date FDA Received11/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention; Disability;
Patient Age55 YR
-
-