Catalog Number 03.812.521
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on an unknown date, during tlif surgery, a cage was inserted using new tpal set equipment.When the surgeon tried to disconnect the insertion device, he was unable to release the grip from the cage.Finally, after multiple attempts, the disconnection was completed, and the surgery was successfully completed.There was a surgical delay of fifteen (15) minutes.This report is for one (1) t-pal advanced applicator inner shaft.This is report 2 of 4 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Narrative (functional issue) couldn't be verified from the provided pictures.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history (lot): part: 03.812.521, lot: l916575, manufacturing site: hägendorf, release to warehouse date: october 10, 2018.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances related to the malfunction were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Additional event information: related product complaint: (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot: part: 03.812.521.Lot: l916575.Manufacturing site: hägendorf.Release to warehouse date: october 10, 2018.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances related to the malfunction were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T-PAL ADVANCED APPLICATOR INNER SHAFT
Type of Device
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key10801737
MDR Text Key215127794
Report Number8030965-2020-08709
Device Sequence Number1
Product Code MAX
UDI-Device Identifier07612334143785
UDI-Public(01)07612334143785
Combination Product (y/n)N
PMA/PMN Number
K181231
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 10/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.812.521
Was Device Available for Evaluation? No
Date Manufacturer Received01/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ADVANCED APPL OUTER SHAFT; APPLICATOR KNOB; UNK - CAGE/SPACERS: T-PAL; ADVANCED APPL OUTER SHAFT; APPLICATOR KNOB; UNK - CAGE/SPACERS: T-PAL
-
-