SYNTHES GMBH BATTERY REAMER/DRILL FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.705
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/26/2020
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Reporter's phone number was not provided.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi: (b)(4).
 
Event Description
It was reported from (b)(6) that the trigger button of the battery reamer/drill device was pressed and did not return to its normal position.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the battery reamer/drill device had low power and the trigger seized in a depressed state, and many components of the device were corroded and damaged.It was further determined that the device failed pretest for check for sticky trigger, and check power with the test bench.Therefore, the reported condition that the trigger button of the device was pressed and did not return to its normal position was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to normal wear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY REAMER/DRILL FOR BPL II
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key10802027
MDR Text Key215130586
Report Number8030965-2020-08419
Device Sequence Number1
Product Code GEY
UDI-Device Identifier07611819491830
UDI-Public7611819491830
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 10/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.705
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2020
Date Manufacturer Received01/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-