Estimated date of implant.The device was not returned for evaluation.The lot history record (lhr) for this product could not be reviewed and a similar complaint query could not be performed because the product was not returned for evaluation and the part and lot numbers were not reported.Based on the information provided, a conclusive cause for the reported stent fracture could not be determined.Fatigue from artery dynamics and motion may have contributed to the stent fracture; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|