Product analysis: one back to stock rfg3 unit received for evaluation.Unit was connected to ac and it performed post successfully with no failures.The reported complaint could not be duplicated.The error log was empty.External inspection found minor cosmetic ware.Internal inspection and evaluation found no discrepancies.The external sd memory card was found to be corrupted and was replaced.The unit passed fan test.Safety test was successfully performed with no failures.Final functional test passed successfully.The unit was placed in burn-in for 12 hours and it passed.The unit performs satisfactorily within medtronic standards.The unit passed burn-in successfully and is ready for packaging.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|