• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND VENA SEAL CLOSURE SYSTEM; AGENT, OCCLUDING, VASCULAR, PERMANENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND VENA SEAL CLOSURE SYSTEM; AGENT, OCCLUDING, VASCULAR, PERMANENT Back to Search Results
Model Number VS-402
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hypersensitivity/Allergic reaction (1907); Ulceration (2116); Ulcer (2274); Injury (2348)
Event Date 10/20/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during procedure, venaseal occluding device was used to treat a segment in the great saphenous vein (gsv) due to venous insufficiency.The lumen was flushed prior to use.Ifu was followed.A guidewire was used for the insertion of the catheter.The procedure was completed normally.Hypersensitivity at post procedure follow up was reported.Patient was place on steroid and benadryl.Medtronic clinician was in touch with attending physician on reported event and it was explained that the dark lesion shown in the photograph is an access site from venaseal treatment in the mid-thigh.Patient is a c6 subject with an active ulcer in the lower leg.Was most likely due to the presence of adhesive in the subcutaneous tissue near the access site.It was explained that physicians would use an 11 blade and explore the area, removing any polymerized adhesive.The reported event does not resemble a straight forward hsr issue, as that does not necessarily involve open lesions.Physician plans to bring back the patient in a couple of days to debride the area and remove any adhesive that may be in or near the access site.No further patient injury reported.
 
Manufacturer Narrative
Date of event updated.Additional information: the catheter tip was located 5cm caudal to the saphenous femoral junction (sfj).Compression of gsv was performed.16 cm of the left great saphenous vein (gsv) were treated with venaseal.The volume of adhesive delivered to the vessel was not recorded.The reaction was noted 6 days post procedure the area was with an active ulcer in the lower leg was debrided 14 days post procedure.The patient health status remains the same.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: the patient is doing fine and pain and symptoms have subsided.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Photo analysis: the customer has provided two photos.Each photo shows the reported access site to the mid-thigh.A dark lesion/ulcer at the access site was observed and suspected adhesive may have been present at the subcutaneous tissue near the access site.The first photo shows dark lesion prior to post-treatment care.The second photo shows the access site diameter of the observed dark lesion was decreased.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENA SEAL CLOSURE SYSTEM
Type of Device
AGENT, OCCLUDING, VASCULAR, PERMANENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key10809034
MDR Text Key215295393
Report Number9612164-2020-04321
Device Sequence Number1
Product Code PJQ
UDI-Device Identifier00643169986268
UDI-Public00643169986268
Combination Product (y/n)N
PMA/PMN Number
P140018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 01/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2022
Device Model NumberVS-402
Device Catalogue NumberVS-402
Device Lot Number58989
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/27/2020
Initial Date FDA Received11/09/2020
Supplement Dates Manufacturer Received11/04/2020
11/12/2020
12/03/2020
01/25/2021
Supplement Dates FDA Received11/12/2020
11/13/2020
12/15/2020
01/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-