RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Failure to Charge (1085)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 09nov2020.
|
|
Event Description
|
During periodic maintenance of the ventilator, the manufacturer¿s international service technician reported a charging failure in the battery.There was no patient involvement.The manufacturer¿s international service technician confirmed the reported issue.The manufacturer¿s international service technician replaced the power management board to address the reported problem.The unit was checked overall, run in tested, cleaned, functionally tested and no abnormality was confirmed.
|
|
Manufacturer Narrative
|
A power management board was return for analysis.A visual inspection of the returned component was performed, no notable conditions were found.An investigation was performed and the product analysis technician reported that no fault was found.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|