SMITH & NEPHEW, INC. GENESIS CALIPER; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 114943 |
Device Problem
Device Markings/Labelling Problem (2911)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/20/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that, during surgery, it was noticed that markers on instrument had become hard to see.Procedure continued with the same device, without a delay or an injury.
|
|
Manufacturer Narrative
|
The device, used in treatment, was returned for evaluation.A visual inspection confirmed the markings are fading on the genesis caliper.The device shows significant signs of wear/usage.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|