It was reported that when the stent delivery system (sds) was opened, the 3.5x23 mm xience sierra stent was loose.The device was not used in the patient.A new, same size xience sierra was used to complete the procedure.There was no reported clinically significant delay in the procedure.No additional information was provided.
|
A visual inspection was performed on the returned stent, and the reported stent dislodgment was confirmed.The stent delivery system (sds) was not returned.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints reported from this lot.Factors that can contribute to stent dislodgement include, but not limited to; improper or inadequate crimping at the time of manufacture, incorrect sheath sizing, positive pressure during preparation, forced sheath removal, handling of the stent during preparation, and interaction with the lesion, accessory devices, and/or previously implanted stents.To ensure this is not the result of a manufacturing issue, all sds are inspected for proper stent placement, stent damage and crimped stent outer diameter.The investigation was unable to determine a conclusive cause for the reported stent dislodgement.It is possible that the technique used to remove the protective sheath or force protective sheath removal resulted in the stent dislodgment from the sds; however, this could not be confirmed.The noted damage on the stent likely occurred during packing for return analysis.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.Na.
|