The device, used in treatment, was returned for evaluation.A visual inspection confirms the tip of the device is broken off.The broken piece was not returned.Also the visual confirms the other end of the device is rounded off.There is also a slight curve/bend in the device.The device shows significant signs of wear/usage.The device was manufactured in 2020.A medical investigation was conducted and this case reports that during the procedure, the cork screw remover was "out of strait." a photo of the returned device shows that there is a slight curve/bend in the instrument.Per email communication, there was no patient injury, and the procedure was completed with a 15 minute delay, using a competitor device.Since no patient harm is alleged, no further clinical assessment is warranted.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of risk management files found that the reported failure was documented appropriately.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
|