• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG VENOVO VENOUS STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG VENOVO VENOUS STENT SYSTEM Back to Search Results
Model Number VENUM14120
Device Problems Misfire (2532); Activation Failure (3270)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/15/2020
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 08/2022).
 
Event Description
It was reported that during stent placement procedure, the stent was partially deployed.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: the lot history record of this lot was reviewed with special attention to the manufacturing and inspection of this product, and the product was found to have met the specification prior to shipment.Investigation summary: the physical sample was not available.One image was provided demonstrating the placed stent inside the vessel.The stent was visibly constricted approximately 10-15mm proximal of the distal end which is the area of the silicone stent brake.Distal and proximal of that section the stent was visible and regularly expanded which indicated that the deployment sheath was retracted and that the stent adhered to the silicone brake, which leads to a confirmed result for expansion issue.A definite root cause for the reported event could not be determined.Labeling review: a review of the relevant instructions for use for this product was conducted.The instructions for use was found to address potential complications and adverse events such as incorrect positioning of the stent requiring further stenting or surgery, intimal injury, and malposition.H10: d4 (expiry date: 08/2022), g4.H11: f10(device), h6 (method, results).H11: section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that during a bilateral stent placement procedure in the iliac veins, the stent allegedly failed to expand as it interacted with the inner catheter.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENOVO VENOUS STENT SYSTEM
Type of Device
VENOUS STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
MDR Report Key10815003
MDR Text Key216225988
Report Number9681442-2020-20028
Device Sequence Number1
Product Code QAN
UDI-Device Identifier00801741103452
UDI-Public(01)00801741103452
Combination Product (y/n)N
PMA/PMN Number
P180037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVENUM14120
Device Catalogue NumberVENUM14120
Device Lot NumberANEV1744
Was Device Available for Evaluation? No
Date Manufacturer Received01/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age65 YR
Patient Weight55
-
-