• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Model Number N/A
Device Problem Power Problem (3010)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/16/2020
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the motor was not running smoothly.The motor and reciprocating arm were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing.It was noted that the device has not been regularly returned for annual pm, per the ifu states: the zimmer air dermatome should be returned every 12 months for inspection and preventative maintenance.Annual factory calibration checks are strongly recommended to verify continued accuracy.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
 
Event Description
It was reported that after connecting air to the device it was not starting up.No further information provided.Investigation showed that the motor was not running smoothly.No adverse event was reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key10817183
MDR Text Key215609201
Report Number0001526350-2020-00937
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 11/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number61576512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2020
Was the Report Sent to FDA? No
Date Manufacturer Received11/09/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/30/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-