Model Number 86-5034 |
Device Problems
Break (1069); Device-Device Incompatibility (2919)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/28/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that on both of the patella cut guides, the metal lift bar that the saw blade goes under to cut has broken off and thus will not hold the saw blade in a position.They were not used on a patient and they will be returned to depuy.No surgical delay.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary examination of the returned device confirmed the reported event.The noted damage is consistent with device use from normal use and servicing and the investigation did not establish a need for corrective action.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.
|
|