Product event summary: the 4fc12 sheath with lot number 0010314891 was returned and analyzed.Visual inspection showed the sheath shaft was kinked and twisted around, 1.86 inches from the tip.The kink prevented smooth steerability when a test balloon catheter was inserted.The sheath distal tip was intact with no fractures, breaks or kinks noted.No teflon delamination was noted at the tip of the sheath shaft.Functional testing of the sheath showed the deflection worked as per specification.In conclusion, the reported kink was confirmed.The sheath failed the returned product inspection due to a shaft kink.If information is provided in the future, a supplemental report will be issued.
|