• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Signal Artifact/Noise (1036); High impedance (1291); Difficult to Insert (1316); Failure to Sense (1559)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2020
Event Type  malfunction  
Manufacturer Narrative
This product is registered as a combination product.
 
Event Description
It was reported that during an implant, no sensing, high, out of range impedance, and noise were observed.All connections checked.It was noted that the is4 adaptor was not fitting lead well, and a new one was provided.However, unable to test the lead.The left ventricular lead was replaced with no issues.There were no adverse health consequences to the patient.
 
Manufacturer Narrative
The complete lead was returned with the associated is-4 psa connector sleeve for analysis with the reported events of no sensing and high lead impedance due to the is-4 adapter not fitting the lead well.Electrical tests did not find shorts or discontinues on any of the conduction paths.The lead connector passed insertion testing into a test device but did not pass insertion testing into the returned is-4 connector sleeve.Dimensional analysis of the connector identified an over-sized diameter in a section of the connector boot that may have contributed to the reported field events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
MDR Report Key10823079
MDR Text Key215946989
Report Number2938836-2020-09214
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2023
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000099625
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2020
Was the Report Sent to FDA? No
Date Manufacturer Received02/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age59 YR
Patient Weight110
-
-