Model Number 26926
Device Problems Premature Activation (1484); Difficult to Advance (2920)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/26/2020
Event Type  Injury  
Event Description
It was reported the stent was prematurely deployed.A contralateral approach was used to access the 95% stenosed, severely calcified and non-tortuous lesion within the superficial femoral artery (sfa).Pre-dilation was performed.A 5 x 150 x 130 innova was advanced over a 0.035 guidewire.The innova did not track well over the guidewire.Due to severe calcification, the stent was deployed above the desired location and the stent did not deploy uniformly.Another innova stent was used to complete the procedure.Post-dilation was performed.There were no patient complications and the patient was stable post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOVA VASCULAR
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key10823149
MDR Text Key215885632
Report Number2134265-2020-15705
Device Sequence Number1
Product Code NIP
UDI-Device Identifier08714729873952
UDI-Public08714729873952
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140028
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number26926
Device Catalogue Number26926
Device Lot Number0026095917
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/27/2020
Initial Date FDA Received11/11/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/29/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1