• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME Back to Search Results
Model Number N/A
Device Problems Break (1069); Intermittent Loss of Power (4016)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 10/14/2020
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under cmp-(b)(4).Once an evaluation of this device is completed, a follow-up/final report will be submitted.Telephone number: (b)(6).
 
Event Description
It was reported that the device is losing power during use, possibly due to a partially broken cable.It could also be a motor fault.No adverse events have been reported as a result of this malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once an evaluation of this device is completed, a follow-up/final report will be submitted.This medwatch is being filed to relay additional information.The following sections have been updated: b3, b4, b5, d10, e2, e3, g4, g7, h2, h3, h6, h10.
 
Event Description
It was reported that the handset was losing power during use.The handset has a partially broken cable, but it could also be a motor fault.Event occurred during surgery and there was no procedure delay nor harm/injury to any persons involved.
 
Manufacturer Narrative
This report is being submitted to report additional information.Review of the most recent repair record determined the device would not operate and motor and switch were replaced and resolved the reported issue.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No new additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
MDR Report Key10824102
MDR Text Key215945012
Report Number0001526350-2020-00942
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup,Followup
Report Date 01/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number63557858
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2020
Was the Report Sent to FDA? No
Date Manufacturer Received01/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-