C.R. BARD, INC. (COVINGTON) -1018233 BARDEX I.C. TEMPERATURE SENSING FOLEY CATHETER Back to Search Results
Model Number 129416M
Device Problems Inaccurate Flow Rate (1249); Insufficient Flow or Under Infusion (2182)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/25/2020
Event Type  Injury  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the foley catheter only drained 50cc of urine after 2 days of working fine.The nurse replaced the foley and 1100cc of urine drained out.
 
Manufacturer Narrative
The reported event was inconclusive.No sample was returned for evaluation.A potential root cause for this failure could be due to "modulus of latex too high".The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use." the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that the bardex foley on the patient worked fine for 2 days later the nurse noticed only 50cc urine drained out during one shift, later nurse tried to manipulate the foley tubing, but no urine drained.The nurse replaced the foley and 1100cc urine drained out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX I.C. TEMPERATURE SENSING FOLEY CATHETER
Type of Device
TEMPERATURE SENSING FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key10826608
MDR Text Key215884701
Report Number1018233-2020-20901
Device Sequence Number1
Product Code MJC
UDI-Device Identifier00801741039775
UDI-Public(01)00801741039775
Combination Product (y/n)N
PMA/PMN Number
K040658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Type of Report Initial,Followup
Report Date 04/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number129416M
Device Catalogue Number129416M
Was Device Available for Evaluation? No
Date Manufacturer Received04/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-