• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. DISPOSABLE ELECTROSURGICAL SNARE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. DISPOSABLE ELECTROSURGICAL SNARE Back to Search Results
Model Number SD-210U-25
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hemorrhage/Bleeding (1888); Perforation (2001)
Event Type  Injury  
Manufacturer Narrative
The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Therefore, the exact cause of the reported event could not be conclusively determined.Since the lot number is unknown, the device history record could not be reviewed.However, omsc has only shipped devices that passed the inspection.In the literature, there was no description of the device's malfunction.
 
Event Description
On october 15, 2020, olympus medical systems corp.(omsc) received literature titled "letters,techniques and images endoscopic shielding technique using polyglycolic acid sheets with fibrin glue for perforation following endoscopic ampullectomy".The purpose was to report a perforation following endoscopic ampullectomy which was successfully treated with filling and shielding by polyglycolic acid (pga) sheets and fibrin glue.In the literature, it was reported that one perforation and bleeding two days after endoscopic ampullectomy was observed.The patient also presented with abdominal pain and, free air was revealed by ct at the same time.The author wrote, "after endoscopic hemoclip treatment, the perforation was filled with pga sheets to protect it from exposure to bile and pancreatic juice.Fibrin glue was then sprayed over the site to allow the pga sheets to adhere to the area.In addition, a nasobiliary drainage (enbd) tube was placed to reduce bile exposure." the esd procedures was performed using an electrosurgical snare (sd-210u-25; olympus).Based on the available information, a direct relationship between the olympus product and the observed adverse event could not be determined.However, according to the figures, the area of a perforation looked the same area of using an electrosurgical snare.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE ELECTROSURGICAL SNARE
Type of Device
DISPOSABLE ELECTROSURGICAL SNARE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10827555
MDR Text Key216109323
Report Number8010047-2020-08846
Device Sequence Number1
Product Code FDI
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K902735
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type literature
Reporter Occupation Other
Type of Report Initial
Report Date 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/12/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Model NumberSD-210U-25
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/15/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age88 YR
-
-