Catalog Number SEPX-8-6-40-135 |
Device Problem
Activation, Positioning or Separation Problem (2906)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/09/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Physician intended to use a protégé rx carotid stent during treatment of a 40mm lesion in the patient¿s internal carotid artery.Lesion exhibited 80% stenosis.Artery diameter reported as 6mm.There was no damage noted to the product packaging prior to use.No issues were noted when removing the device from the packaging.Ifu was followed and the device was prepped without issue.Deployment issues are noted.There were not issues noted to the deployment mechanisms or device handle prior to deployment issue.It is reported the device was delivered to the lesion and was attempted to be deployed.One third of the stent was deployed but could not be fully deployed (partial deployment).The device was replaced to complete the procedure.No patient injury reported.
|
|
Manufacturer Narrative
|
Additional information: the stent was exposed when removed from the patient.The whole device was removed from the patient without intervention and was replaced to complete the procedure.No vessel damage was noted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Device evolution the device was returned with the touhy-borst loose and with biologics visible in the device.The device was returned with approximately 6mm of the distal end of the stent exposed during functional testing the device could not be flushed and a 0.014¿ guidewire could not be loaded.The device was soaked in warm water overnight and flushed with tergazyme in an attempt to clear the biologics, but this was unsuccessful.The device was loaded into a deployment fixture and the stent was deployed with a maximum peak force of 1.451lbs (less than 3.0lbs).The stent and sds inner guidewire lumen assembly were examined; no abnormalities were noted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|