• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GLOBUS MEDICAL, INC. CREO; CREO THREADED LOCKING CAP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GLOBUS MEDICAL, INC. CREO; CREO THREADED LOCKING CAP Back to Search Results
Model Number 1119.0010
Device Problem Loose or Intermittent Connection (1371)
Patient Problem Insufficient Information (4580)
Event Date 10/02/2020
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for evaluation as it was retained by the hospital.Imaging provided shows the locking caps at l2 on both sides of the construct had come loose.It appears that the rod had come out of the screw head on the right side of l2.The exact cause of the reported issue could not be determined.
 
Event Description
It was reported that a revision surgery was done to replace the locking caps and screw heads at l2, which had come loose post-operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CREO
Type of Device
CREO THREADED LOCKING CAP
Manufacturer (Section D)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon 19403
Manufacturer (Section G)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon 19403
Manufacturer Contact
daniel paul
2560 general armistead avenue
audubon 19403
6109301800
MDR Report Key10830736
MDR Text Key217443065
Report Number3004142400-2020-00147
Device Sequence Number1
Product Code MNH
UDI-Device Identifier00889095059496
UDI-Public00889095059496
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K124058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1119.0010
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/15/2020
Initial Date FDA Received11/12/2020
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age72 YR
-
-