Catalog Number SPT-080000S
Device Problem Device-Device Incompatibility (2919)
Patient Problems Foreign Body Reaction (1868); Inflammation (1932)
Event Date 10/20/2020
Event Type  Injury  
Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.Voluntary recall ra 2012-067 was initiated for abgii and rejuvenate modular stems and necks due to the potential risks associated with these devices.The reported corrosion is considered to be under the scope of this recall.No further investigation is required.
 
Event Description
It was reported that the patient's right hip was revised.No patient complaints were reported to the rep.Intra-operatively, minor corrosion was observed at the stem/ neck junction.The rejuvenate modular stem and femoral head were revised.There are no allegations against the femoral head.Rep confirmed that no further information will be released by the hospital or surgeon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REJUVENATE STRGHT PRFIT TMZF MOD STEM SIZE 8
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
diana rogers
210 centennial avenue
centennial park, elstree
borehamwood WD6 3-SJ
UK   WD6 3SJ
2082386500
MDR Report Key10831829
MDR Text Key216136567
Report Number0002249697-2020-02428
Device Sequence Number1
Product Code MEH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092561
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial
Report Date 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/12/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2016
Device Catalogue NumberSPT-080000S
Device Lot NumberMJT488
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/20/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/03/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-2090-2012
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age74 YR
Patient Weight116
-
-