(b)(4).Since no device has been received for analysis, no product investigation can be performed, and the customer complaint cannot be confirmed.Manufacturing record evaluation (mre) cannot be conducted because no lot number was provided by the customer.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Biosense webster manufacturer's reference number (b)(4) has two complaints that are related to the same incident.The hemostatic valve separation malfunction reported under 2029046-2020-01697.Manufacturer's reference number: (b)(4).
|
It was reported that a male patient underwent cardiac ablation procedure for atrial fibrillation (afib) with thermocool® smart touch® sf bi-directional navigation catheter and suffered cardiac tamponade requiring surgical intervention.In addition, a hemostatic valve separation issue occurred with the carto vizigo¿ 8.5f bi-directional guiding sheath ¿ medium.During the case the hemostatic valve on the carto vizigo¿ 8.5f bi-directional guiding sheath ¿ medium was leaking dark blood.The sheath was replaced, and the issue resolved.The event did not require percutaneous or surgical removal no other intervention to stop the bleeding.The hemodynamics was not compromised (several ml blood loss).It was also reported that a pericardial effusion was noticed as the patient's blood pressure dropped.The pericardial effusion was confirmed by ice.Caller reported that the medical intervention provided was a medication (ibutilide 10 mg and protamine 45 ml).The effusion got smaller and the pressure went up.The patient was reported to be in stable condition and the pressure is 109/ 66.Effusion was discovered during use of biosense webster products.Prior to effusion ablation of the cavotricuspid isthmus line was performed but the event occurred during mapping phase.Surgery was required immediately after the procedure was aborted.The final outcome is unknown.The caller did not have information about the need for extended hospital stay.They could only tell that the patient was set to stay overnight.Transseptal puncture was performed with baylis nrg needle.There was no evidence of steam pop.The irrigation flow was set to 2ml/min.No error messages were observed on biosense webster equipment during the procedure.Graph, dashboard, vector and visitag modules were used for force visualization.The visitag stability settings were set to 3 mm for 3 sec, fot25% at 3g.Respiration was used as additional visitag filter and tag index as prospective color option.The physician opinion is that the cause of the event was difficult transseptal puncture performed on the patient.Since the cardiac tamponade may be life threatening, it was assessed as mdr reportable under the thermocool® smart touch® sf bi-directional navigation catheter.The issue with the hemostatic valve was assessed as a mdr reportable malfunction hemostatic valve separation issue under the carto vizigo¿ 8.5f bi-directional guiding sheath ¿ medium.
|