Additional manufacturing narrative: other, other text: the returned device was evaluated.The touchscreen was not responsive on some areas of the screen due to contamination found on the touchscreen flex cable.During functional testing, the device was able to obtain measurements.No product performance issue identified related to spo2 and pulse rate.A service history record review reveals that this unit was in the field for over eleven (11) months with no previous reported issues related to this reported event.Initial reporter zip code exceeded the maximum allowable characters, zip code is as follows: nw1 6jl, corrected data: b3: date of event updated from "(b)(6) 1901" to "(b)(6) 2020".
|