Model Number 72200419
Device Problem Difficult to Advance (2920)
Patient Problem No Information (3190)
Event Date 10/21/2020
Event Type  Injury  
Event Description
It was reported that, during an arthroscopic surgery, the monofilament could not feed through the rollers of the "accu-pass (70°, upbend, sterile)".The procedure was successfully completed without delay using a competitor device.No additional complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10 h3,h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCU-PASS SUTURE SHUTTLE 70 DEGREE
Type of Device
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key10835227
MDR Text Key216085316
Report Number1219602-2020-01831
Device Sequence Number1
Product Code LXH
UDI-Device Identifier03596010542397
UDI-Public03596010542397
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 04/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/06/2022
Device Model Number72200419
Device Catalogue Number72200419
Device Lot Number2042711
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/21/2020
Initial Date FDA Received11/13/2020
Supplement Dates Manufacturer Received04/02/2021
Supplement Dates FDA Received04/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-