• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Model Number EL5ML
Device Problems Failure to Advance (2524); Failure to Fire (2610)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/08/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4); batch #: u94p56.Investigation summary: the analysis results found that the el5ml device was returned with no apparent damage.In an attempt to replicate the reported event, the device was tested for functionality.Upon testing, the device was fired, and the clips did not advance into the jaw.The device was noted to have the tip of the advancer bent.The instrument was disassembled and upon disassembly, the advancer was confirmed to be bent, and eleven clips were found inside clip track.The reported complaint was confirmed.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Please reference the instruction for use for more information.Please note that prior to loading a clip in the jaws, ensure that the demarcation between the jaws, and the device shaft is past the end of the trocar cannula.Additionally, excessively applying a side load to the jaws, causing them to partially collapse, could result in a clip malformation.The device jaws should be fully open, and parallel upon initiating the firing of the device.It should be noted that as part of our quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Event Description
It was reported that during a lap cholecystectomy, when squeezing clip applier, it was not loading clips.Clips were jammed in the device.They opened another clip applier that worked fine to complete case.No patient complications.
 
Manufacturer Narrative
(b)(4).Date sent: 11/13/2020.H6: device codes: failure to advance is being added, 2524.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key10835713
MDR Text Key216135307
Report Number3005075853-2020-05971
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 10/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberU94P56
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2020
Initial Date Manufacturer Received 10/26/2020
Initial Date FDA Received11/13/2020
Supplement Dates Manufacturer Received10/26/2020
Supplement Dates FDA Received11/13/2020
Patient Sequence Number1
-
-