Generator analysis was completed on the returned device.Results of diagnostic testing indicated the device was operating properly and communicated properly.Electrical test results showed that the pulse generator performed according to functional specifications.The pa worksheets were reviewed and no anomalies were noted.There were no adverse functional, mechanical, or visual issues identified with the returned generator.The lead was explanted and returned due to reported pain and product analysis was completed.Note that since a portion of the lead (including) the electrode array portion was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.An abraded opening was identified in the outer and the inner silicone tubing of the negative coil.The lead assembly has dried remnants of what appear to have once been body fluids inside the inner and the outer silicone tubing.No obvious point of entrance was noted other than the identified tubing openings.The lead fracture allegation was not duplicated in the pa lab.Other than the above mentioned observations and typical wear and explant related observations, no other anomalies were identified in the returned lead portions.
|