H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of scanner keeps freezing within 10 seconds every time it is turned on was confirmed; the probe would freeze when scanning after 5 minutes.The scanner would still function normally, but could not scan.The root cause of the reported failure was identified as the beamformer board would cause the scanning to freeze.A history review of serial number dyczac503 showed no other similar product complaint(s) from this serial number.H3 other text : evaluation findings are in section h.11.
|