• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO GASTROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO GASTROSCOPE Back to Search Results
Model Number EG-2990I
Device Problem Device Damaged by Another Device (2915)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Pentax medical has not received any further information for this event and therefore, considers this medwatch report closed.(b)(4).
 
Event Description
Pentax medical was made aware of an event that occurred at a facility in the united states.On (b)(6) 2019, pentax received a service notification repair request for "resistance auxiliary biopsy channel".During inspection the following was found: "accessory stuck in primary operation channel", involving the pentax medical video gastroscope, model eg-2990i, serial number (b)(4).Pentax medical customer service team performed good faith efforts to good to hear hows circumstance of when the event occurred.The user stated this occurred during reprocessing and the endoscope was subsequently removed from circulation and called in for service.On (b)(6) 2019, repairs were completed where the stuck accessory was removed, and miscellaneous repairs were performed and returned to the user.On (b)(6) 2020, a device history record(dhr) review for model eg-2990i, serial number (b)(4)was performed under ivai-20-110014, the dhr review confirmed the endoscope was manufactured on (b)(6) 2007 under normal conditions, passed all required inspections, and was released accordingly.Also, there were no reworks or concessions and the dates of approval for shipment and actual date shipped were confirmed.Instructions for use(ifu), includes the following warning section 2-1-3, 3) "after using operational/cleaning accessories (e.G., forceps, needles, snares, brushes etc.) with the endoscope, carefully check that all accessories are intact and that no parts have fallen off and become lodged within the endoscope's instrument/suction channel.Furthermore, ensure that any therapeutic devices (e.G., clips, stents, etc.) passed through the channel are accounted for after use.On (b)(6) 2016, pentax issued a u.S.Urgent field correction which is an ifu addendum for endoscopes with instrument channels.This addendum covers any operational/cleaning accessories and therapeutic devices which can become lodged in the endoscope's instrument channel.It reminds customers to carefully check that all accessories are intact, that no parts have fallen off and become lodged within the endoscope's instrument/suction channel and to ensure that any therapeutic devices (e.G., clips, stents, balloons, etc.) passed through the instrument channel and are accounted for after use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO GASTROSCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0012
JA 
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0012
JA  
Manufacturer Contact
william goeller (temporary)
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key10837254
MDR Text Key238965570
Report Number9610877-2020-00205
Device Sequence Number1
Product Code FDS
UDI-Device Identifier04961333129492
UDI-Public04961333129492
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131902
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEG-2990I
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/05/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/23/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-