Product event summary: the sheath 4fc12 with lot 0010372188, was returned and analyzed.The sheath was intact with no apparent issues.The performance test with sentinel blackbelt leak tester revealed the pressure decay is in the acceptance range.The shaft and valve are leak tight with no apparent issue.The air aspiration issue was not reproduced.In conclusion, the sheath passed the returned product inspection per specification.If information is provided in the future, a supplemental report will be issued.
|