• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Model Number EL5ML
Device Problem Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/23/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # u94k8m.Investigation summary: the analysis results found that the el5ml device was received with one jaw disengaged from the cam.This condition would not allow the jaws to collapse in order to form the clips.No functional test was performed due the condition of the device.In order to evaluate the condition of the internal components of the device, it was disassembled.Upon disassembling, no anomalies were found.In addition, eight clips were found inside clip track.No conclusion could be reached as on what caused the jaw disengaged.Although there is no direct evidence of use error, the instructions for use do contain the following caution: possible causes for the condition of the jaw disengaged from the cam may be inadvertent force, twisting or pressure being placed on the device jaws, using the jaws of the device as a dissector/ retractor or damage to the jaws while entering the trocar.It should be noted that as part of our quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformance's were identified.
 
Event Description
It was reported that during an unknown procedure, the clip was not fed into the jaws at the first firing.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 00969
*   00969
6107428552
MDR Report Key10840378
MDR Text Key216327283
Report Number3005075853-2020-06007
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 10/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2020
Initial Date Manufacturer Received 10/29/2020
Initial Date FDA Received11/13/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/15/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-