DEPUY SYNTHES PRODUCTS LLC BATTERY HANDPIECE/MODULAR FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 05.001.201 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/01/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
|
|
Event Description
|
It was reported from (b)(6) that the bottom trigger of the battery handpiece/ modular device had broken off.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2020.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the device was fractured ¿ trigger, had a leak tightness test failure, a cracked/damaged housing, and a damaged component.It was further determined that the device failed pretest for check roundness of housing, check for sticky triggers, leakage test using bubble emission technique and general condition.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to component failure from normal wear.
|
|
Search Alerts/Recalls
|
|
|