Model Number N/A |
Device Problem
Material Frayed (1262)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/16/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under cmp-(b)(4).Once an evaluation of this device is completed, a follow-up/final report will be submitted.
|
|
Event Description
|
It was reported that the cable is cut.It occurred after the case was finished on (b)(6)2020.There was no harm or delay, and no alternate contact.No adverse events were reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the cable was cut.The plug harness was replaced and resolved the reported issue.Device is used for treatment.Review of complaint history identified additional similar complaints for the reported item and part and lot combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
|
|
Event Description
|
No additional event info.
|
|