SILK ROAD MEDICAL INC. ENROUTE TRANSCAROTID STENT SYSTEM; ENROUTE SDS Back to Search Results
Model Number SR-0830-CS
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Stroke/CVA (1770); Brain Injury (2219)
Event Date 10/23/2020
Event Type  Injury  
Manufacturer Narrative
The product associated with this complaint was not returned to the manufacturer for analysis.A review of manufacturing records for this device was completed and no issues were identified that could have led to the adverse event reported.Although there is no indication that a malfunction of the srm device occurred, the cause of the post-operative complication is unknown.A follow-up mdr will be submitted if additional information is obtained.Silk road medical will continue to monitor for occurrences of similar events.
 
Event Description
It was reported that after the completion of a transcarotid artery revascularization (tcar) procedure, the physician stated that the patient because hypotensive (~50-60 systolic) in the recovery room.After treating the hypotension, she became hypertensive (~200-225 systolic).The physician noted that the patient experienced a stroke.The patient was taken back to surgery to have stent removed.The physician removed black ulcerated plaque with white specks of calcium from the stent.The patient experienced slurred speech, facial droop, and left side weakness.The patient was returned to the operating room for re-intervention and post-dilation was performed.The patient was returned to the icu.The following day, the patients facial droop and leg mobility improved, however the left arm weakness remained.The physician noted that there were blood pressure issues with this patient that may have contributed to the patient having a stroke and subsequently having the stent removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENROUTE TRANSCAROTID STENT SYSTEM
Type of Device
ENROUTE SDS
Manufacturer (Section D)
SILK ROAD MEDICAL INC.
1213 innsbruck drive
sunnyvale CA 94089
Manufacturer (Section G)
SILK ROAD MEDICAL INC.
1213 innsbruck drive
sunnyvale CA 94089
Manufacturer Contact
magaly boardman
1213 innsbruck drive
sunnyvale, CA 94089
4087309002
MDR Report Key10846139
MDR Text Key216662835
Report Number3014526664-2020-00111
Device Sequence Number1
Product Code NIM
UDI-Device Identifier00811311020461
UDI-Public(01)00811311020461(17)210630(10)17878438
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/30/2021
Device Model NumberSR-0830-CS
Device Catalogue NumberSR-0830-CS
Device Lot Number17878438
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/23/2020
Initial Date FDA Received11/16/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/30/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Disability;
Patient Age76 YR
-
-