Model Number 74018901
Device Problem Break (1069)
Patient Problems No Consequences Or Impact To Patient (2199); No Code Available (3191); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/22/2020
Event Type  Injury  
Event Description
It was reported that during tka procedure the impactor chipped during impaction.It is unknown if there was a surgical delay nor how was the procedure completed.
 
Manufacturer Narrative
H3, h6: the affected complaint device, used in treatment, was not returned for evaluation.Therefore, a product analysis could not be performed and complaint could not be confirmed.The clinical investigation concluded that this case reports the tibial impactor chipped during impaction.Per complaint details, there was no patient injury reported.It is unknown whether a backup device was required, or if there was a surgical delay.Therefore, since no patient harm is alleged, no further clinical assessment is warranted at this time.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved or product information, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY TIBIAL IMPL IMPACTOR
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED,
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10849624
MDR Text Key216555828
Report Number1020279-2020-06511
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010557285
UDI-Public03596010557285
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 08/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number74018901
Device Catalogue Number74018901
Device Lot NumberUNKNOWN
Initial Date Manufacturer Received 10/22/2020
Initial Date FDA Received11/16/2020
Supplement Dates Manufacturer Received08/05/2021
Supplement Dates FDA Received08/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-