Catalog Number 1011-01-101
Device Problem Material Fragmentation (1261)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi (b)(4).
 
Event Description
It was reported that the male end of the cup impactor device handle sheared off at the coupling.It was not reported if the event occurred during a surgical procedure.It was not reported if there was a delay to a planned procedure.It was not reported if there was a spare device available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.Quality engineering evaluated the impactor device and the reported condition that the male end of the cup impactor handle sheared off at the coupling was confirmed.An assessment was performed on the cup-adapter t-handle and found the proximal end of the cloverleaf broke off from the t-handle, consistent with having been dropping or other radial (off-axis) force, which is improper handling.It was further determined that the device failed pretest for visual assessment.The assignable root cause of this condition was determined to be traced to user, which is user error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KINCISE PINNACLE SHELL/LINER IMPACTOR
Type of Device
IMPACTOR, ADAPTERS, BATTERY, BATTERY CHARGER
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
MDR Report Key10854409
MDR Text Key222540215
Report Number1045834-2020-01905
Device Sequence Number1
Product Code GEY
UDI-Device Identifier00850915006099
UDI-Public850915006099
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 11/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1011-01-101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2020
Date Manufacturer Received12/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-