DEPUY SYNTHES PRODUCTS LLC KINCISE PINNACLE SHELL/LINER IMPACTOR; IMPACTOR, ADAPTERS, BATTERY, BATTERY CHARGER
|
Back to Search Results |
|
Catalog Number 1011-01-101 |
Device Problem
Material Fragmentation (1261)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi (b)(4).
|
|
Event Description
|
It was reported that the male end of the cup impactor device handle sheared off at the coupling.It was not reported if the event occurred during a surgical procedure.It was not reported if there was a delay to a planned procedure.It was not reported if there was a spare device available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.Quality engineering evaluated the impactor device and the reported condition that the male end of the cup impactor handle sheared off at the coupling was confirmed.An assessment was performed on the cup-adapter t-handle and found the proximal end of the cloverleaf broke off from the t-handle, consistent with having been dropping or other radial (off-axis) force, which is improper handling.It was further determined that the device failed pretest for visual assessment.The assignable root cause of this condition was determined to be traced to user, which is user error.
|
|
Search Alerts/Recalls
|
|
|