Model Number EL5ML |
Device Problem
Failure to Form Staple (2579)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/03/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Batch #: unknown.The lot/batch was not provided; therefore, a manufacturing record evaluation could not be performed.Attempts are being made to retrieve the device.To date, the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
|
Event Description
|
It was reported that during a general surgery procedure, upon discharge and attaching of surgical clip, it was noted clip scissored upon itself.Attempted to remove malformed staple and case was successfully completed with no patient consequence.
|
|
Manufacturer Narrative
|
(b)(4).Date sent: 12/1/2020.D4: batch # u94a0x.F10.Component code: others (g07).H6: component code: others (g07).Investigation summary the analysis results found that the el5ml device was returned with no apparent damage.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed nine conforming clips.The event described could not be confirmed as the device performed without any difficulties noted.Although there is no direct evidence of use error, the instructions for use do contain the following: "caution: do not excessively twist or torque the instrument jaws when positioning or firing the instrument on a tubular structure or vessel.Excessive twisting or torqueing may result in clip malformation." it should be noted that as part of our quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
|
Search Alerts/Recalls
|
|
|