• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. MIC-KEY GASTRIC-JEJUNAL FEEDING TUBE KIT - 22 FR, 3.0 CM - 45 - NON-ENFIT; DH EF BALLOON TUBES PRODUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. MIC-KEY GASTRIC-JEJUNAL FEEDING TUBE KIT - 22 FR, 3.0 CM - 45 - NON-ENFIT; DH EF BALLOON TUBES PRODUCTS Back to Search Results
Model Number 0270-22-3.0-45
Device Problem Material Puncture/Hole (1504)
Patient Problem Vomiting (2144)
Event Date 07/27/2020
Event Type  malfunction  
Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is not possible as no lot number was provided.All information reasonably known as of 18-nov-2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).
 
Event Description
It was reported one device had a small hole just below the balloon so when feeding in jejunal port it came out g-port.Additional information received 27-oct-2020 stated the patient was vomiting with tube feedings.A kidneys, urethra and bladder (kub) x-ray noted the device was in proper placement.The device was replaced.The removed device noted a clogged small pin hole below the balloon on the jejunal side of the device.
 
Manufacturer Narrative
The investigation concluded that without a sample functional evaluation could not be performed in order to confirm or duplicate the reported incident.A root cause could not be determined.All information reasonably known as of 18-dec-2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIC-KEY GASTRIC-JEJUNAL FEEDING TUBE KIT - 22 FR, 3.0 CM - 45 - NON-ENFIT
Type of Device
DH EF BALLOON TUBES PRODUCTS
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
MDR Report Key10864184
MDR Text Key222019880
Report Number9611594-2020-00229
Device Sequence Number1
Product Code KNT
UDI-Device Identifier00350770956756
UDI-Public00350770956756
Combination Product (y/n)N
PMA/PMN Number
K926581
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0270-22-3.0-45
Device Catalogue NumberN/A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/24/2020
Patient Sequence Number1
Patient Age27 YR
-
-