The field service engineer (fse) conducted a site visit and upon arrival, the fse removed the cover and found the cup transfer mechanism was stopped over the incubator.The fse was able to confirm the reported error via the error logs.The fse was able to reproduce the error by performing a cup transfer test, after a few times it failed.The fse replaced the cup transfer assembly along with the incubator and performed all the alignments.The fse performed a cup transfer test several times and all was completed with no errors.The fse inspected and adjusted all temperatures.The fse ran quality control (qc) and verified instrument operation.The aia-900 analyzer is functioning as expected.No further action required by field service.A 13-month complaint and service history review for serial number (b)(4) from (b)(6) 2019 through aware date (b)(6) 2020 was performed for similar complaints.There were two similar complaints including this one identified during the search period.The aia-900 operator's manual under section 12 - flags and error messages states the following: c.Transfer cup pickup failure cause: the cup sensor s063 failed to detect the cup after the cup was grasped.Action: please contact the tosoh local representatives.Check s063, the cup pickup position, and the cup pickup operation.The most probable cause of the reported event is faulty cup transfer assembly.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|