Model Number 1458Q/86 |
Device Problems
Difficult to Insert (1316); Failure to Advance (2524)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/05/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This product is registered as a combination product.The results, method, and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
During an implant procedure, the stylet was unable to be inserted into the left ventricular (lv) lead.The lv lead was explanted and replaced to resolve the event and the patient was stable.
|
|
Manufacturer Narrative
|
The reported event of stylet failure to advance was confirmed.As received, a complete lead was returned in one piece.Visual inspection and x-ray examination found the inner coil kinked at the s-curve region.A stylet could not be fully inserted into the lead due to the inner coil being kinked at the s-curve region.The cause of stylet failure to advance was due to kinked coil which is consistent with procedural damage.Electrical testing did not find any indication of conductor fractures or internal shorts.
|
|