Model Number 000724
Device Problems Break (1069); Fluid/Blood Leak (1250); Migration (4003)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/16/2020
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 06/2023).
 
Event Description
It was reported that approximately thirty-five days post gastrostomy tube placement, the tube allegedly dislodged and fell out of the patient.It was revealed that the balloon ruptured.There was no reported patient injury.
 
Event Description
It was reported that approximately thirty-five days post gastrostomy tube placement, the tube allegedly dislodged and fell out of the patient.It was revealed that the balloon ruptured.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: a complaint history review was performed.This is the seventh complaint reported for this product/lot number combination.The device history records were reviewed and there was nothing found to indicate there was a manufacturing related cause for these reports.Investigation summary: the device was not returned for evaluation.The investigation is inconclusive for the reported tube dislodged, rupture and break issues.A definitive root cause could not be determined.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE, 24F
Type of Device
FEEDING DEVICE
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key10873166
MDR Text Key217462847
Report Number3006260740-2020-20588
Device Sequence Number1
Product Code KNT
UDI-Device Identifier00801741037177
UDI-Public(01)00801741037177
Combination Product (y/n)N
PMA/PMN Number
K063118
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number000724
Device Catalogue Number000724
Device Lot NumberNGDW1641
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/26/2020
Initial Date FDA Received11/20/2020
Supplement Dates Manufacturer Received06/08/2021
Supplement Dates FDA Received06/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age47 YR