• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SENSATION SHORT THROW; SNARE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SENSATION SHORT THROW; SNARE, FLEXIBLE Back to Search Results
Model Number M00562693
Device Problems Break (1069); Entrapment of Device (1212); Retraction Problem (1536)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/04/2020
Event Type  Injury  
Manufacturer Narrative
The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that a sensation short throw was used during a colonoscopy procedure performed on (b)(6) 2020 according to the complainant, during the procedure, the snare handle mechanism broke and the snare failed to close properly during the polypectomy.Reportedly, the snare loop became stuck in the polyp and the tech was forced to cut the snare wire by the handle in order to force pull the loop into the catheter.The snare was successfully removed from the patient using suction.The procedure was completed with a different captivator snare.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be ok.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSATION SHORT THROW
Type of Device
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752 1242
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752-1242
5086834015
MDR Report Key10875620
MDR Text Key217345633
Report Number3005099803-2020-05533
Device Sequence Number1
Product Code FDI
UDI-Device Identifier08714729501671
UDI-Public08714729501671
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131700
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/09/2022
Device Model NumberM00562693
Device Catalogue Number6269-40
Device Lot Number0024909216
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/05/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/10/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age53 YR
-
-