SMITH & NEPHEW, INC. REPL BIT FOR TORQUE WRENCH; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71565657 |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/04/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that, during a tka surgery, the repl bit for torque wrench was found worn.S&n backup available.No surgical delay or injury to patient reported due this event.
|
|
Manufacturer Narrative
|
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The hex end of the device is dull and worn rendering it inoperable.The device shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|