The device was not returned for evaluation, therefore a no product return engineering evaluation was performed.During the manufacturing process, the device was visually inspected and tested several times.All the inspections are conducted on 100% of units, except in the case of product verification (pv) testing, where the tested units are chosen on a sampling basis.All tested sample units for this lot passed pv testing.These inspections and tests during the manufacturing process support that it is unlikely that a non-conformance contributed to the reported complaint.A review of edwards lifesciences risk management documentation was performed for this case. the reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of the failure mode is not required at this time. the complaint for central regurgitation was unable to be confirmed due to the unavailability of the device/imagery for evaluation.A review of the dhr and lot history, revealed no indication that a manufacturing non-conformance contributed to the complaint.Additionally, a review of manufacturing mitigations supports that proper inspections are in place during the manufacturing process to detect issues relating to the complaint.Review of ifu/training manuals revealed no deficiencies.During the manufacturing process, all sapien 3 ultra valves are 100% visually inspected for defects and 100% tested for coaptation prior to release for distribution.This makes it highly unlikely that a manufacturing defect or device malfunction would contribute to the event.Per the instructions for use (ifu), central regurgitation is a potential adverse event associated with bioprosthetic heart valves and the transcatheter aortic valve replacement (tavr) procedure.There are multiple patient and procedural factors that alone or in combination can cause or contribute to central regurgitation including malposition of the valve, impingement of a leaflet due to the guide wire, over inflation of the deployment balloon, post dilation of the implanted valve, and slow recovery of adequate ventricular flow post valve deployment and rapid pacing. all of these factors have the potential to contribute to suboptimal coaptation of the valve leaflets and cause central aortic insufficiency.As reported, the valve was deployed with an extra volume (+1cc) and ¿post-dilatation was performed in case there was a piece of the patient anatomy protruding into the cage or on top of the leaflet.The ai only got worse.¿ per training manual, ¿the delivery system requires a prescribed volume for thv deployment and proper function¿ and ¿if regurgitation is central rather than paravalvular, do not post-dilate.¿ overexpansion may have initially led to the improper coaptation of valve leaflet, resulting in central leakage.The post dilation may have further damaged the leaflet, worsening the central leakage.As such, available information suggests that procedural factors (over inflation of balloon and post dilation) may have contributed to the complaint event.However, a definitive root cause was unable to be determined at this time.No labeling/ifu deficiencies were identified; therefore, no corrective or preventative action, nor product risk assessment is required at this time. complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.
|