A field service engineering (fse) was at the customer's site to address reported event.Fse confirmed reported error by reviewing the error log and reproduced error by running a sample.Fse observed that the error occurred when the rack moves few inches away from home.Fse inspected alignment of the x1 pitch sensor flag and found it was misaligned.Fse adjusted and aligned the flag, performed rack rotation test run, performed a daily check run and performed a quality control run; all results passed without error and within acceptable range.No further action required by field service.The aia-900 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no similar complaints identified during the search period.The aia-900 operator's manual under section 12 - flags and error messages states: [2300] s.Loader step feed failure.Cause: the pitch sensor s071 failed to go on after the step feed.Action: check to see if there is any impediment to the sample rack feed.If the trouble reoccurs, contact the tosoh local representatives.Check s071 and the step feed mechanism.The most probable cause of the reported event was due to misaligned sample loader x1 pitch sensor.
|