Catalog Number 74122542
Device Problem Compatibility Problem (2960)
Patient Problems Pain (1994); Metal Related Pathology (4530)
Event Date 11/29/2010
Event Type  Injury  
Manufacturer Narrative
Complaint reference: (b)(4).
 
Event Description
It was reported that the patient underwent revision surgery due to pain associated to a possible metal related pathology.An r3 tha system was revised during this procedure.
 
Manufacturer Narrative
H3, h6: it was reported that left hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.The hemi head, sleeve and r3 liner are no longer sold by smith & nephew.A review of the complaint history for the r3 shell, r3 liner, hemi head, modular sleeve, screw and stem was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the shell, liner, stem and head.A similar complaint has been identified for the screw.This will continue to be monitored.A similar complaint has been identified for the sleeve.However, as the device is no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Review of manufacturing records did not reveal any waivers, concessions, manufacturing or material abnormalities that could have contributed to this issue.Review of the product instructions for use found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.The available medical documents were reviewed.The clinical information provided of pain, brownish clear fluid and thickened pseudocapsule may be consistent with a reaction to metal debris.However, the source and the root cause cannot be determined with the available documentation.It cannot be concluded that the reported clinical findings are associated with the implant failure.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMI HEAD 42MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
MDR Report Key10882215
MDR Text Key217647501
Report Number3005975929-2020-00454
Device Sequence Number1
Product Code NXT
UDI-Device Identifier00885556070710
UDI-Public885556070710
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 04/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/30/2013
Device Catalogue Number74122542
Device Lot Number08HW18403
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/28/2020
Initial Date FDA Received11/22/2020
Supplement Dates Manufacturer Received04/21/2021
Supplement Dates FDA Received04/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-